- Rating
- 5
- Clarity
- 4
- Confidence
- 5
- Originality
- 3
- Quality
- 3
- Significance
- 3
Read review
Summary
This work aims to improve depth-wise visualization in reflectance confocal microscopy (RCM), where sparse axial sampling and anisotropic resolution limit the interpretation of tissue structure. The authors propose CD-RCM, a feedforward transformer-based framework that formulates continuous-depth reconstruction as a novel-view synthesis problem and generates intermediate optical sections at arbitrary depths from sparsely sampled RCM stacks. To account for the unique imaging geometry of RCM, the method models axial acquisition using virtual camera representations and incorporates a skin-specific perceptual feature loss derived from a domain-adapted self-supervised backbone. It is relatively speaking, a novel application of neural view synthesis to optical sectioning microscopy and demonstrates its potential for continuous-depth reconstruction in RCM.
Strengths and weaknesses
The paper addresses an important and underexplored problem in biomedical imaging: improving depth-wise visualization of reflectance confocal microscopy (RCM) stacks through continuous-depth reconstruction. The motivation is compelling, as RCM volumes are highly anisotropic and difficult to interpret in cross-sectional views, despite their widespread use in dermatology. The authors propose CD-RCM, a feedforward transformer-based framework that formulates axial densification as a novel-view synthesis problem and generates intermediate optical sections at arbitrary depths. The work further introduces a skin-specific perceptual loss derived from a domain-adapted self-supervised feature extractor to better capture the microstructural characteristics of RCM images.
A key strength of the paper is its domain-specific framing. Rather than directly applying existing neural rendering methods, the authors carefully motivate why RCM differs from both conventional multi-view imaging and tomographic reconstruction, and adapt the model design to reflect the acquisition geometry of optical sectioning microscopy. The problem itself is clinically relevant, and the proposed feedforward inference framework is computationally attractive compared to optimization-based neural rendering approaches. I also appreciate that the study is conducted on a substantial in vivo RCM dataset collected by the authors, which is valuable given the scarcity of publicly available benchmarks in this modality.
The paper is generally well written and easy to follow. The imaging motivation, architectural design, and training procedure are clearly described, and the qualitative visualizations effectively illustrate the intended application of isotropic reconstruction and arbitrary-plane sectioning.
However, I have several concerns regarding the experimental evaluation and the strength of the conclusions. Most importantly, the comparison is limited to classical interpolation methods such as B-spline, cubic spline, and Gaussian interpolation. Since CD-RCM is a learned transformer-based model trained specifically for this task, outperforming traditional interpolation approaches is expected. The paper does not compare against modern learning-based interpolation, reconstruction, or view-synthesis methods, making it difficult to determine whether the gains arise from the proposed RCM-specific design or simply from the use of a large learned model.
A second concern is the limited scope of validation. The dataset consists of healthy skin imaged using a single acquisition platform, and all experiments are performed within this setting. While collecting such a dataset is a notable effort, the current evaluation provides limited evidence regarding generalization across devices, institutions, patient populations, or pathological conditions. As a result, some of the broader claims regarding generalizable reconstruction should be interpreted cautiously.
I am also concerned about the biological validity of synthesized slices. The paper frequently refers to anatomically plausible intermediate sections and highlights improved cross-sectional visualization. However, no clinician evaluation, biological validation, or uncertainty analysis is provided. Because the model generates unobserved intermediate structures, it is important to assess whether diagnostically relevant features are faithfully reconstructed rather than hallucinated. This issue is particularly important if the method is intended to support downstream clinical interpretation.
Finally, although the proposed skin-specific perceptual loss appears beneficial, its contribution is demonstrated primarily through image-quality metrics derived from the same dataset. Additional evidence linking these improvements to biologically meaningful reconstruction quality would strengthen the paper.
It proposed a new problem and the microscopy motivation is compelling. The adaptation of neural view synthesis concepts to optical sectioning microscopy is interesting and potentially impactful. However, the experimental evaluation does not yet provide sufficiently strong evidence to support the broader claims of generalizable and anatomically faithful reconstruction. Stronger baselines and more rigorous biological validation would substantially strengthen the work.
Questions
- Can the authors compare CD-RCM against at least one modern learning-based interpolation or reconstruction baseline?
Currently, the evaluation is limited to classical interpolation methods (B-spline, cubic spline, and Gaussian interpolation). While these are reasonable references, it is difficult to assess the value of the proposed architecture without comparison to a learned baseline.
- How robust is CD-RCM across different imaging settings and tissue types?
The dataset collected by the authors is a valuable contribution given the scarcity of publicly available RCM data. However, all experiments appear to be performed on healthy skin acquired using a single imaging platform. Could the authors discuss or provide evidence regarding expected performance on different skin types, anatomical sites, devices, or pathological conditions?
3, How can we assess whether synthesized intermediate slices are anatomically faithful rather than plausible-looking hallucinations?
The paper demonstrates improvements in image-quality metrics and visual appearance, but clinical utility depends on preserving diagnostically relevant structures. Can the authors provide additional analysis, uncertainty estimates, or expert assessment regarding the fidelity of reconstructed intermediate slices?
- Will the dataset and code be publicly released?
The dataset appears to represent a substantial effort and addresses a modality with relatively limited public resources. Releasing the dataset, train/test split, and preprocessing pipeline would significantly improve reproducibility and increase the broader impact of the work. If public release is not possible due to consent or institutional constraints, could the authors clarify whether qualified academic access will be provided and discuss any restrictions on data availability?
- How sensitive is the method to residual registration errors?
The paper notes that stack alignment remains a limitation. Since RCM data can contain motion artifacts and imperfect registration, it would be useful to understand how reconstruction quality changes under varying levels of misalignment and whether the model can be made more robust to these effects.
Limitations
No. The authors appropriately acknowledge residual registration errors and the limited size and diversity of the dataset. However, several important limitations are not sufficiently discussed. First, the evaluation only compares against classical interpolation methods, making it difficult to assess the advantages of the proposed approach relative to modern learning-based alternatives. Second, the paper does not address the possibility that synthesized intermediate slices may contain anatomically plausible but incorrect structures, which is particularly important if the method is intended to support clinical interpretation. Finally, the current evaluation is limited to healthy skin acquired on a single imaging platform, and the implications for generalization to pathological cases, other devices, or broader clinical settings should be discussed more explicitly.
Paper formatting concerns
NA
Final justification
The authors have addressed my concerns, and since the paper now falls under the User-Inspired category, I would revise my recommendation to accept. My primary motivation is that the work is interesting, timely, and has the potential to inspire future research in the field. As microscopy datasets are increasingly being integrated with state-of-the-art machine learning models, this study addresses an emerging and relevant direction. Furthermore, the authors demonstrate the potential advantages of their method, and given that the novelty requirement is less stringent for the User-Inspired category, I believe the paper meets the bar for an acceptance.
We thank the reviewer for the careful assessment and for recognizing CD-RCM’s domain-specific framing as a novel application of neural view synthesis to optical sectioning microscopy. This is the central objective of the paper: to establish continuous-depth RCM reconstruction as a technically viable novel-view synthesis problem and motivate further research at this intersection.
We thank the reviewer for recognizing the “domain-specific framing” of CD-RCM as “relatively speaking, a novel application of neural view synthesis to optical sectioning microscopy and demonstrates its potential for continuous-depth reconstruction in RCM.” This is indeed the central objective of this paper, i.e., to establish continuous-depth RCM reconstruction as a technically viable novel-view synthesis problem, and to motivate further research on this intersection.
A. It is difficult to determine whether the gains arise from the proposed RCM-specific design or simply from the use of a large learned model… Compare CD-RCM against at least one modern learning-based interpolation or reconstruction baseline.
We agree that this distinction should be made clearer and report additional learned baselines in our response to Reviewer 82mA (point B) and Reviewer BWXy (point A) to directly examine whether generic neural model capacity alone is sufficient for this task. The NeRF-like per-stack coordinate MLP underperforms every classical interpolation baseline, despite accurately fitting the observed input slices. Because it is optimized independently for each stack, it cannot learn a cross-stack prior over depth-dependent tissue morphology. Similarly, FILM does not reliably preserve RCM morphology: zero-shot inference introduces discontinuous and morphologically implausible structures, while fine-tuning improves PSNR and SSIM primarily by producing smoother predictions, at the expense of both perceptual metrics. In addition, our training-objective ablation holds the CD-RCM architecture fixed and shows that replacing the proposed skin-specific perceptual supervision with a standard perceptual loss degrades performance across all metrics.
Taken together, these results indicate that the gains do not arise merely from applying a large learned model. Rather, they depend on the RCM-specific formulation of CD-RCM, which leverages the structured axial acquisition process of optical sectioning microscopy, represents depth as a continuous and queryable dimension, and learns a cross-stack prior over depth-dependent tissue morphology. We will revise the manuscript to make this distinction and the supporting evidence more explicit.
B. Generalization across devices, institutions, patient populations, pathological conditions, imaging settings, and tissue types; biological validity of synthesized slices.
We agree that these broader forms of generalization and biological validity have not yet been established. In the current paper, generalization refers specifically to inference on held-out RCM stacks and skin sites within the evaluated data distribution (i.e., healthy skin acquired using the same imaging platform). It does not imply demonstrated generalization across devices, institutions, patient populations, pathological conditions, or other tissue types. We will revise the manuscript to define this scope more precisely and avoid broader interpretations of the term “generalizable.”
This limited scope is intentional for the present proof-of-concept study. The primary goal is to demonstrate that continuous-depth RCM reconstruction can be formulated as a novel-view synthesis problem and to provide initial evidence that this direction is technically viable. Establishing this feasibility is a necessary first step before undertaking larger studies involving pathological tissue, more diverse populations, multiple devices and institutions, and task-specific clinical annotations.
As described in our response to Reviewer 82mA (point C), we will also include an evaluation facilitated by clinical experts to assess the anatomical fidelity of the reconstructed views, identify morphologically implausible or hallucinated structures, and examine their potential utility for DEJ localization. We emphasize that this evaluation will provide an initial assessment of reconstruction fidelity and potential clinical utility, rather than establish clinical efficacy or broad biological validity. We will explicitly identify external validation across pathological tissue, imaging devices, institutions, and patient populations as necessary future work.
C. Public release of the code and dataset.
We plan to release the model code and trained checkpoints to facilitate reproducibility and further research. Unfortunately, we cannot publicly release the RCM dataset because of institutional and data-governance constraints associated with the clinical data. We will add a clear data-availability statement to the manuscript. Subject to the required institutional approvals and data-use agreements, we will also work with our clinical collaborators to provide controlled access to qualified academic researchers where possible.
D. Sensitivity to residual registration errors.
We thank the reviewer for this valuable suggestion. CD-RCM assumes that the input slices are approximately laterally aligned. Although SIFT-based affine registration substantially reduces inter-slice motion, residual misalignment remains possible and may degrade reconstruction quality.
To quantify this sensitivity, we will add a controlled robustness analysis with increasing levels of synthetic registration error applied to the conditioning slices using small two-dimensional affine perturbations, and report the resulting impact on the quantitative metrics. We will also evaluate registration-noise augmentation during training, using random translations and other small affine perturbations, to determine whether it improves robustness to residual misalignment.
We will report these results in the camera-ready version, and clarify that CD-RCM requires reasonably aligned input stacks, and substantial motion remains an important limitation and direction for future work.