Summary
The paper proposes MiCo, a physics-informed conditional flow-matching framework for angular super-resolution (ASR) in diffusion-weighted MRI: recovering a dense q-space signal from a sparse angular acquisition to shorten scan time. The method exploits the linearity of the DKI cumulant-expansion forward model (y = Hx) through two closed-form components. (1) A DTI-based source initialization replaces the standard Gaussian prior of flow matching: a 6-parameter diffusion-tensor fit to the sparse observations synthesizes a dense estimate, so the network only models the residual. (2) A microstructure-consistency loss penalizes prediction error only in the range subspace of the DKI design matrix H (orthogonal projection R = H(H^T H)^{-1} H^T), shown to be exactly equivalent to a Fisher-metric discrepancy in DKI parameter space and implemented with no per-step parameter fit. Experiments on HCP-YA and UK Biobank evaluate signal NMSE, downstream DTI/DKI scalar maps (FA, MD, MK, ...), and FOD reconstruction. The central empirical claim is a dissociation: signal-level NMSE saturates across modern learned methods, yet MiCo attains lower error on downstream microstructural parameters, with ablations attributing signal gains to the initialization and parameter gains to the consistency loss.
Strengths and weaknesses
Strengths This paper has a clinically relevant core insight distinguishing signal-space fidelity from microstructure fidelity, focusing on clinically used downstream metrics (FA/MD/MK maps, FODs) instead of raw signal MSE as a meaningful conceptual contribution. It rigorously applies the forward model using closed-form DKI linearity, with an elegant derivation showing Fisher-metric discrepancy simplifies to a range-space projection, naturally deriving the 1/(1-t)^2 weighting; its lightweight, architecture-agnostic loss is reusable for all linear forward models. Its DTI-initialized residual flow design is physically justified, cutting signal-NMSE by 34% on HCP-YA versus noise initialization. Experiments are thorough, covering two large public datasets, three sparsity levels, three evaluation dimensions, multiple classical and learned baselines, region-wise analysis, and ablation tests separating its two components via Pareto weight tuning. The writing is modest and transparent: it admits saturated signal NMSE with MiCo matching DISCUS in signal performance, highlights downstream divergence as its main merit, and frankly discusses limitations including DKI model constraints, healthy-only cohorts, and ODE inference cost.
Weaknesses W1. The headline “dissociation” claim is currently under-substantiated. The central HCP-YA downstream result (Fig. 3) is presented only as relative-error curves and a scatter, with no companion numeric table of absolute FA/MD/MK errors with std comparable to the UKB Table 5. Because the effect sizes are small (e.g., the consistency loss tightens MK by ~2.8% on HCP-YA), and several gains over DISCUS fall within the 3-seed variability (signal-NMSE cells are traded in Table 1; FOD/ACC is best in only 5/9 cells and trails MAP-MRI/SHORE at higher sparsity), the reader cannot judge statistical significance. A full HCP-YA numeric table plus paired significance testing across subjects/seeds is needed to firmly establish the main claim. W2. The ablation behavior is dataset-dependent and only explained post-hoc. On UKB the DTI source alone degrades WM-MK by +12.5% (Table 2), and only adding the consistency loss recovers and surpasses the baseline. The regime explanation (b_max = 2000 vs 3000, cumulant-truncation validity) is plausible but is not directly tested; a restriction to voxels where the DKI fit is well-conditioned, or a b^3-residual analysis, would make the 'two complementary contributions' narrative more convincing. W3. The individual components are incremental relative to the cited literature. Data-dependent / residual initialization for diffusion-flow bridges (I2SB, ResShift, UPMRI) and projection onto the forward operator's range for data consistency (range-null decompositions) are both established. The contribution is the specific, exact instantiation for the DKI operator plus the evaluation framing; the conceptual delta over UPMRI (range-space projection for MRI flow matching) in particular should be stated explicitly. W4. Scope limits keep the clinical claim aspirational. MiCo operates per voxel with no spatial neighborhood, whereas dMRI ASR typically benefits from spatial coherence; and all data are healthy young/adult cohorts, while the motivating applications (Alzheimer's, stroke, tumor) are exactly the pathological, out-of-distribution regime where the DKI prior is least valid. Both are acknowledged, but they bound the practical impact demonstrated.
Questions
Q1. Please provide a full numeric table (mean+/-std over the 3 seeds) of HCP-YA downstream errors (FA, MD, MK, ideally AK/RK) for all methods, mirroring the UKB Table 5. Are MiCo's improvements over DISCUS statistically significant under a paired test across subjects/seeds? Q2. On UKB the DTI-source-only variant degrades WM-MK (+12.5%) before the consistency loss recovers it. Can you support the cumulant-validity explanation directly — e.g., restrict evaluation to voxels where the DKI fit is well-conditioned, or quantify the b^3 residual — rather than via the b_max heuristic? Q3. What is the conceptual and empirical difference between your range-space DKI loss and (a) UPMRI's projection onto the parallel-MRI forward range and (b) classical range-null / data-consistency layers? An explicit positioning, or a head-to-head, would clarify novelty. Q4. At 10% sparsity the DTI init is fit to only ~5-10 directions. How often is H_obs ill-conditioned, and what is the failure mode of y_init in that case? Q5. Inference uses a 25-step midpoint ODE per voxel. What is the wall-clock cost for a full brain volume versus DISCUS and an analytic fit, and how does few-step distillation affect downstream accuracy? Q6. Is the qualitative crossing-fiber advantage (Fig. 5) reflected in any quantitative tractography/connectivity metric, or only in ACC and a single qualitative figure?
Limitations
Yes — limitations are adequately and honestly discussed in Appendix A: dependence on DKI model validity (free water/CSF/demyelination), healthy-only narrow-demographic cohorts with unverified generalization to pediatric/geriatric/pathological populations, scope limited to angular (not spatial) super-resolution, and multi-step ODE inference cost. The gaps I would add are the absence of any pathological/out-of-distribution evaluation despite the clinical framing, and the lack of statistical-significance analysis for the modest effect sizes. No negative societal concerns beyond standard medical-imaging considerations.
Paper formatting concerns
The central HCP-YA downstream result (Fig. 3) has no companion numeric table, which weakens verifiability of the main claim. Separately, the injected text hidden in the page-2 and page-26 footer margins of the PDF text layer (see the Integrity Note) is both a formatting anomaly and an integrity issue and must be removed.
Final justification
After considering the responses in the rebuttal and the other reviews I retain my original rating of the paper.
Thank you for the detailed review. We address the requested evidence for distinguishing signal from microstructure fidelity below.
[Q1,W1] Full HCP-YA results and statistical comparisons
10% context
20% context
30% context
Values are pooled voxel-median absolute errors (×10−2). MiCo is lowest for FA/MK/AK/RK at every context, while CFM or I²SB is lower for MD/AD/RD. We will report all seven metrics and narrow the claim accordingly.
On matched q-space samples, masks, and 40,960 voxels, two-sided paired Wilcoxon tests give:
All 15 comparisons favor MiCo (Δ<0 favors MiCo), while its signal NMSE is 4.48–5.31% higher. Thus, MiCo better preserves the fitted microstructural quantities without minimizing signal error.
HCP-YA component ablation at 20% context:
The NMSE change is negligible despite statistical detectability from the large voxel count; the DKI loss reduces MK error. Mean±SD over three runs separately reports run-to-run variability.
[Q2,W2] Direct test of DKI model validity
We directly measure mismatch between the dense ground-truth signal and DKI:
rDKI(y)=∥y∥2∥(I−ΠR)y∥2,
where $\Pi_R$ projects onto the rank-22 DKI subspace. On UKB at 20% context, the median is 0.052 and p95 is 0.165. Stratifying test voxels by this residual:
The DKI loss helps only in the low-residual half, so component complementarity is conditional on DKI fit. We will replace the bmax explanation with this analysis, report the high-residual failure regime, and narrow the claim.
[Q3,W3] Difference from UPMRI and range–null/data-consistency methods
The projection algebra is shared; our claim concerns the operator and its role. UPMRI/range–null methods use an acquisition operator to enforce measured k-space consistency. MiCo uses the DKI parameter-to-dense-q-space operator H, whose range is the DKI signal subspace, not the angular-subsampling range.
During training, this projection defines a supervised loss equivalent to the Fisher-metric discrepancy between DKI estimates:
∥x^−x∥H⊤H2=∥ΠR(y^−y)∥22.
The standard flow-matching loss supervises the orthogonal component, and inference is not hard-projected. We will add this operator-level comparison and narrow the novelty claim. A UPMRI numerical comparison is not task-aligned because it addresses parallel-MRI k-space reconstruction with a different operator and domain.
[Q4] Conditioning of the DTI initialization
The context set pools all nonzero-b measurements and retains all b=0 observations; DTI is fitted once across shells. Thus, the relevant sample size is the total context count, not the per-shell count.
θ^=(AC⊤AC+ΛC)−1AC⊤yC,
where AC∈RNC×7 is the restricted design matrix, yC contains log-signals, and ΛC is the diagonal Tikhonov regularizer. For gi=(gix,giy,giz), row i is
[−bigix2,−2bigixgiy,−bigiy2,−2bigixgiz,−2bigiygiz,−bigiz2,−1].
We will correct “ordinary least squares” to “Tikhonov-regularized least squares.”
The evaluated joint systems are not severely ill-conditioned; this does not establish robustness to substantially sparser, shell-imbalanced, or pathological inputs.
[Q5] Runtime and few-step inference
We benchmarked in-memory HCP-YA batches of 40,960 voxels on one H100 NVL, including transfer, preprocessing, and terminal-signal generation but excluding loading, disk output, and downstream fitting.
At 20% context, 25→5 steps gives a 5.9× speedup; HCP-YA NMSE/FA/MK change by +4.3%/+1.9%/+2.3%, and UKB by +4.4%/+0.8%/+9.9%. Five steps is a practical fixed-checkpoint setting, not distillation or evidence that fewer steps are universally better. DISCUS remains ~9× faster, so speed is not claimed.
[Q6,W4] Angular evidence and scope
We will move a compact FOD-ACC table to the main results. The corrected summary is that MiCo is best on all three shells at 10% context, best in 5/9 shell-context cells overall, and within one ACC point of the leader in every cell; MAP-MRI and SHORE lead several higher-context cells. This supports competitive, not uniformly superior, angular fidelity.
We will also add the requested companion to Fig. 5 with voxels colored by their ACC in the updated manuscript. The review interface does not permit us to attach that figure or an external link during discussion.
The paper does not evaluate tractography, connectivity, pathology, or spatial context. We will remove language implying those benefits and limit the empirical claim to voxel-wise signal, scalar-microstructure, and FOD reconstruction in healthy HCP-YA and UKB cohorts. Pathology-specific and tractography validation remain future work rather than evidence assumed by the present study.
[Formatting/integrity concern] Hidden PDF text
The footer text was neither authored nor inserted by us, it originates from the conference PDF-processing pipeline.